Estimation of
Voriconazole in Biological Matrix by HPTLC-MS
Santosh R. Tambe1*, Sanjay D. Sawant2
1Assistant Professor, M. G. V’s Pharmacy
College, Panchavati, Nashik
2Professor and Principal, STES SKN
College of Pharmacy, Kondhwa (BK), Pune – 411 048
*Corresponding
Author E-mail: santoshsrt.mgv@gmail.com
ABSTRACT:
A High Performance Thin
Layer Chromatography Tandem Mass Spectrometry (HPTLC-MS) method was developed
for estimation of Voriconazole in a biological matrix. HPTLC-MS is a rapid and
economic technique for appropriate identification and quantitation of active
pharmaceutical ingredients in pharmaceutical dosage forms and also biological matrices. The Rf value of the drug was 0.33 ± 0.011
using Toluene: Ethyl Acetate: Methanol: (6:0.5:2.5 v/v/v) as the mobile phase.
Linearity was obtained within a concentration range of 100 – 600 ng/band with
regression coefficient 0.999. The
accuracy of the proposed method for analyzing the API in terms of percentage
recovery was 100.1 with percentage RSD 1.53%, both of which were within the
specified limits, verifying that the method is accurate and precise. LOD and
LOQ were 20. 47 ng/band and 62.05 ng/band, respectively. The recovery after
application to biological matrix was 97.7± 1.46 % proving its applicability to
biological analysis. Specificity study showed that the components of biological
matrix did not interfere with the analyte, thereby confirming the specificity
of the method. Mass spectrometric analysis of the samples resulted in a
spectrum showing molecular ion peak at 350.1 m/z. HPTLC-MS method resulted in
direct confirmation in identifying the drug. .
This HPTLC method seems to be convenient as well as less time consuming
and thus illustrating its wide applicability.
KEY WORDS: HPTLC-MS, Planer Chromatography, Mass spectrometry, Voriconazole.
INTRODUCTION:
Voriconazole (VOR)
(Fig. 1), a derivative of Fluconazole, is a second-generation triazole
antifungal agent. VOR is chemically (2R, 3S)-2-(2,
4-difluorophenyl)-3-(5-fluoro-4-pyrimidinyl)-1-(1H-1, 2,
4-triazol-1-yl)-2-butanol. The primary mode of action of VOR is the inhibition
of fungal cytochrome P-450-mediated 14-α-lanosterol demethylation, an
essential step in fungal ergosterol biosynthesis. The accumulation of 14
alpha-methyl sterols correlates with the subsequent loss of ergosterol in the
fungal cell wall and this may be responsible for the antifungal activity of VOR1.
It has been shown to be more selective for fungal cytochrome P-450 enzymes than
for various mammalian cytochrome P-450 enzyme systems1, 2.
VOR is used to treat
serious fungal infections such as invasive aspergillosis (a fungal infection
that begins in the lungs and spreads through the bloodstream to other organs)
and esophageal candidiasis (infection by a yeast-like fungus that may cause
white patching in the mouth and throat). It works by slowing the growth of the
fungi that causes infection3, 4.
Figure 1- Structure of
VOR
Literature survey
revealed number of methods available for estimation of VOR which includes UV
Spectrophotometric methods5-7, High Performance Liquid
Chromatographic methods8-14, Stability-indicating HPLC method15,
Experimental design approach in HPLC method16, along with some
modification in the methods such as RP-HPLC17,18. Also some advanced
methods like LC-MS/MS19-21, UPLC method22, Stability
indicating UPLC method23 and Ultra-Performance Liquid
Chromatography–Tandem Mass Spectrometry Method24 are reported.
The present work is
based on the development of High Performance Thin Layer Chromatography Tandem Mass Spectrometry
(HPTLC-MS). Thin
layer chromatography (TLC) is a simple, cost-effective, and easy-to-operate,
planar chromatographic technique that has been used routinely in general
chemistry laboratories since several decades to separate chemical and
biochemical compounds. Traditionally, chemical and optical methods are employed
to visualize the analyte spots on the TLC plate
However, direct identification and structural characterization of the
analytes on the TLC plate are not possible through these methods. Hence, there
has been long held interest in the development of interfaces that allow TLC to
be combined with mass spectrometry (MS); one of the most efficient analytical
tool for structural elucidation25. This method is very rapid as well
as economical as compared to other sophisticated techniques.
MATERIALS AND METHODS:
Apparatus
Camag HPTLC system
comprising of Camag Linomat V semiautomatic sample applicator, Camag
TLC Scanner 3, Camag twin-trough developing chamber (10 X 10 cm), UV
cabinet, Camag Win CATS software, and Hamilton syringe (100 µL), were
used in the present research work.
Reagents and Materials
Silica Gel 60 F254
TLC plates (10 X 10 cm, layer thickness 0.2 mm, E. Merck, Darmstadt, Germany)
were used as stationary phase. Methanol, Chloroform and Toluene (AR grade,
Fisher Scientific, India) were used for mobile phase preparation.
Preparation of
standard and sample solution
Accurately weighed 10
mg of drug was transferred to 10 mL volumetric flask and diluted to 10 mL using
methanol to get concentration 1 mg/mL. From this stock solution, 1 mL was
further diluted to 10 mL using methanol.
Preparation of matrix
for analysis
For the analysis of
the drug from a biological matrix, a simulated fluid was prepared in the
laboratory; the compositions of which is given below in Table 1
Chromatographic
conditions
The experiment was
performed on silica gel 60 F254 aluminum sheets (10 X 10 cm) as
stationary phase, using mobile phase comprised of Toluene: Ethyl Acetate:
Methanol (6:0.5: 2.5v/v/v). TLC plates were prewashed with methanol and
activated in an oven at 110 şC for 10 min prior to chromatographic experiment.
The solutions were applied on TLC plate in the form of bands of 6 mm width
under a stream of nitrogen gas using a Camag Linomat V semiautomatic
sample applicator. Ascending development to 70 mm was performed in 10 cm x 10
cm Camag twin trough glass chamber saturated with vapors of mobile phase
for 15 min. The developed TLC plate was air dried and then scanned from 200 to
400 nm using Camag TLC scanner 3 using WinCATS software. The drug
showed reasonably good response at 253 nm keeping the slit dimension of 6.00 x
0.30 mm and scanning speed of 20 mm/s.
Table 1- Master
formula for biological matrix
Ingredients |
Quantity (g) |
Glucose |
2.2065 |
Lactic Acid |
0.5 |
Acetic Acid |
0.013 |
Albumin |
0.2164 |
Urea |
1.0432 |
Sodium chloride |
0.2621 |
Potassium chloride |
0.2485 |
Calcium Chloride |
0.026 |
Glycerol |
0.1 |
Distilled water |
up to 250 mL |
Linearity and calibration
graph
Linearity was obtained
over the concentration range 100 – 600 ng/band by applying 1 – 6 µL from
standard solution (100ng/µL) as a band of 6 mm width on the TLC plate. All the
chromatographic conditions were maintained as above. The Rf value for
VOR was 0.33 ± 0.011 along with asymmetry factor 1.2 (Fig 2). Peak areas of VOR
were plotted against corresponding concentrations and least square regression
analysis was performed to generate the calibration equation (Fig 3).
Figure 2- Densitogram of standard VOR (200 ng/band, Rf = 0.33 ±
0.011)
Table 2- Estimation of
VOR in biological matrix
Concentration (ng/band) |
Area (µV. Sec) (Average, n = 6) |
Mean % Recovery* ± SD |
% RSD* |
200 |
2670.70 |
97.7 ± 1.46 |
1.49 |
Average of six
determinations
Figure 3- Calibration graph of VOR
Application of the
proposed method for estimation of VOR in biological matrix
Appropriate volume (4
mL) from stock solution (1 mg/mL) was withdrawn in 10 mL volumetric flask and
volume was adjusted up to mark with methanol. This standard solution having
concentration 400 µg/mL was used for estimation of VOR in biological matrix.
Spiked samples (100
ng/μL, 6 replicate) were prepared by adding 0.25 mL solution from above
standard solution of VOR (400 µg/mL) to 0.25 mL of prepared biological matrix
and making volume to 1 mL with methanol. The contents of the tubes were
vortexed for 3 min and then centrifuged for 10 minutes at 2500 rpm. After
centrifugation, 2 µL supernatant aliquots of each replicate were applied as
bands on the plate and the plate was developed as per the standard conditions.
Results are shown in Table 2.
Mass Spectrometric conditions
The triple quadrupole
system was an API 4000 Q TRAP (ABSCIEX, CA, USA) LC-MS/MS spectrometer
fitted with an electro-spray ionization interface. The ESI-MS was operated in
both positive and negative detection mode. Calibration of the mass analyzer was
performed by infusion (10µL min-1) of a commercial standard of
polypropylene glycol which was supplied by AB SCIEX using a 1mL Hamilton
syringe and monitored mass-to-charge ratios (m/z) in the 59-1800 mass
range. The ESI source conditions were: ion spray voltage, 5500 V; nebulizer gas
(GS1), 50 psi; curtain gas, 25 psi; turbo gas (GS2), 50 psi; collision gas
(CAD), 7 psi and ion source temperature 470 °C. The drug detection and mass
identification were performed. Q1 and Q3 quadrupoles were set to unit
resolution. The mass spectrum of VOR was displayed as reported in Fig. 4.
Validation of HPTLC
method:
Validation of the
proposed method was carried out with various parameters such as precision,
accuracy, specificity, LOD, LOQ.
·
Accuracy and Precision:
Intra and inter day
accuracy and precision evaluations were performed by repeated analysis of VOR
in biological matrix. The run consisted of a three replicates of each 200, 300,
400 ng/band of VOR. The overall precision of the method expressed as relative
standard deviation and recovery of the method. Intraday accuracy ranged from
101.69 % to 98.61 % and precision was 1.57 %. Inter day accuracy ranged from
101.92 % to 99.12 % and precision was 1.50 %. The mean Recovery, standard
deviation (SD), coefficient of variation (% RSD) was evaluated and their
results were tabulated in Table 3.
Figure 4- Full scan mass spectrum of VOR
Table 3- Results of
intra-day and inter-day precision and accuracy
Concentration (ng/band) |
Intra-day precision |
Inter-day precision |
||||||
Area (µV. Sec) |
% Recovery ± SD |
Mean % Recovery* ±
SD |
% RSD* |
Area (µV. Sec) |
% Recovery ± SD |
Mean % Recovery* ±
SD |
% RSD* |
|
200 |
2741.5 |
101.69 ± 0.84 |
99.99 ± 1.57 |
1.57 |
2743.5 |
101.92 ± 0.89 |
100.21 ± 1.50 |
1.50 |
200 |
2724.9 |
2728.9 |
||||||
200 |
2754.2 |
2760.2 |
||||||
300 |
3590.2 |
99.66 ± 0.59 |
3579.2 |
99.58 ± 0.65 |
||||
300 |
3560.1 |
3557.1 |
||||||
300 |
3582.7 |
3590.7 |
||||||
400 |
4396.5 |
98.61 ± 0.69 |
4399.5 |
99.12 ± 0.92 |
||||
400 |
4404.1 |
4464.1 |
||||||
400 |
4441.7 |
4431.7 |
Figure 5- Densitogram
of Blank sample of matrix
·
Specificity:
Specificity is the ability to measure accurately and
specifically the analyte of interest in the presence of other components that
may be expected to be present in the sample matrix. The specificity of the
proposed method is illustrated in Fig. 5 where, a blank sample of matrix was
spotted on the plate. The densitogram shows that there were no endogenous peaks
at the retention time of VOR. The components of biological matrix did not
interfere with the analyte, thereby confirming the specificity of the
analytical method.
·
LOD and LOQ:
The limit of detection
and limit of Quantitation were calculated for VOR by using the standard deviation
and slope obtained from linearity studies and it was found to be LOD 20.47 ng /
band and LOQ 62.05 ng / band.
RESULTS AND DISCUSSION:
The reported method is
based on the principle of chromatographic separation and spectrometric
identification of VOR in the biological matrix. VOR showed Rf value
0.33 ± 0.011 along with asymmetry factor 1.2 using Toluene: Ethyl Acetate:
Methanol (6:0.5:2.5 v/v/v) as mobile phase. Linearity was obtained within
concentration range 100 – 600 ng/band with regression coefficient 0.999. Sample
recovery was found to be 97.7 ± 1.4 % when the method was applied to biological
matrix. Mass spectroscopy was resulted in the spectra showing molecular ion
peak at 350.1 m/z indicating mass of VOR.
The developed HPTLC
method was validated by performing various parameters in which the accuracy and
precision was evaluated based on percentage recovery and percentage relative
standard deviation. The mean percentage recovery was found to be 99.99 ± 1.57 %
with mean percentage RSD 1.57 % for intraday accuracy and precision. Similarly,
mean percentage recovery 100.21 ± 1.50 % along with mean percentage RSD 1.50 %
was obtained for interday accuracy and precision. Specificity was carried out
by spotting blank matrix which resulted in clear densitogram without any peak
in the region where VOR showed a peak. The results observed in specificity
study showed that the components of biological matrix did not interfere with
the analyte, thereby confirming the specificity of the analytical method. The
sensitivity of the method was determined by calculating LOD and LOQ for VOR
which was found to be 20.47 ng/band and 62.05 ng/band, respectively.
CONCLUSION:
Development and
validation of HPTLC method for the analysis of Voriconazole in the biological
matrix have been reported. The method was accurate, reproducible, specific and
applicable for biological analysis. Additionally, the HPTLC-MS, a less time
consuming method, for identification the drug was also developed to
authenticate the analyte.
ACKNOWLEDGEMENT:
The authors would like
to thank Management, Principal and all the staff members of M.G.V’s Pharmacy
College, Panchavati, Nashik, and Dr. Santosh Gandhi, S. S. P. M. S. College of
Pharmacy, Pune, for their valuable support and sincere help in the present research
work.
REFERENCES:
1.
Martindale: The Complete Drug Reference. 36th edition,
Pharmaceutical Press, Lambeth High Street, London. 2009; 550-551.
2.
Budavari S, editor.
The Merck Index. 13th ed. White House Station (NJ): Merck and Co.
Inc; 2001. p. 1789.
3.
Hiroyasu K. et al. Short-Term Therapy with
Luliconazole, a Novel Topical Antifungal Imidazole, in Guinea Pig Models of
Tinea Corporis and Tinea Pedis. Antimicrobial
Agents and Chemotherapy. 56; 2012: 3136-3143.
4.
Goodwin ML and Drew RH. Antifungal Serum Concentration
Monitoring: An Update. Journal of
Antimicrobial Chemotherapy. 61; 2008: 17–25
5.
Bonazzi D. et al. Determination of Imidazole
Antimycotics in Creams by Supercritical Fluid Extraction and Derivative UV
Spectroscopy. Journal of Pharmaceutical
and Biomedical Analysis. 18; 1998: 235–240.
6.
Roy CP, Rao KS and Annapurna MM. Estimation of
Voriconazole in Pharmaceutical Dosage forms by Zero and First Order Derivative
Spectrophotometric Methods. International
Journal of Pharmacy Technology. 2 (3); 2010: 837-846.
7.
Roy S. et al. Development and Validation of New
Analytical Method for Voriconazole by Using UV spectrophotometer. International Journal of Pharmacy &
Technology. 3 (1); 2011: 1904-1912.
8.
Emilio CB. et al. Development and Validation of a
Fast HPLC/Photodiode Array Detection Method for the Measurement of Voriconazole
in Human Serum Samples. A Reference Laboratory Experience. Enferm Infecc Microbiol Clin. 31; 2013: 23-8.
9.
Cheng S, Qiu F, et al. Development and Validation
of a Simple and Rapid HPLC Method for the Quantitative Determination of
Voriconazole in Rat and Beagle Dog Plasma. Journal of Chromatographic Science .45 (7); 2007: 409-14.
10.
Oates. et al. Determination of Voriconazole
Related Compound F in Voriconazole Using IC. Thermo Fisher Scientific, Sunnyvale,
CA, USA. Available from: URL:
http://www.dionex.com/en-us/webdocs/113759-AN1022-IC-Voriconazole-Cmpnd-F-25Sep2012-AN70191_E-0912S.pdf
11.
Gage R., Stopher D. A. A Rapid HPLC Assay for
Voriconazole in Human Plasma. Journal
of Pharmaceutical and Biomedical Analysis. 17; 1998: 1449–1453.
12.
Perea S. et al. Comparison of High-Performance
Liquid Chromatographic and Microbiological Methods for Determination of
Voriconazole Levels in Plasma. Antimicrobial
Agents and Chemotherapy. 44 (5); 2000: 1209-1213.
13.
Pennick G.J. et al. Development and Validation of
a High-Performance Liquid Chromatography Assay for Voriconazole an Analytical
method for the Determination of Voriconazole. Antimicrobial Agents and Chemotherapy. 47 (7); 2003: 2348-2350.
14.
Khoschsorur G. A. et al. Isocratic
High-Performance Liquid Chromatographic Method with Ultraviolet Detection for
Simultaneous Determination of Levels of Voriconazole and Itraconazole and its
Hydroxy Metabolite in Human Serum. Antimicrobial
Agents and Chemotherapy. 2005: 3569–3571.
15.
Ping G., Yuru L. Development and Validation of a
Stability-Indicating HPLC Method for Determination of Voriconazole and its
Related Substances. Journal of
Chromatographic Science. 47; 2009: 594-598.
16.
Srinubabu G. et al. Development and Validation of
a HPLC Method for the Determination of Voriconazole in Pharmaceutical
Formulation Using an Experimental Design. Talanta. 71 (3); 2007: 1424-1429.
17.
Bharathi J. et al. Validated RP-HPLC Method for
the Estimation of Voriconazole in Bulk and Tablet Dosage Form. International Journal of Research in
Pharmaceutical and Biomedical Sciences. 1 (1); 2010: 14-18.
18.
Sadasivudu P., Shastri N., Sadanandam M. Development and
Validation of RP-HPLC and UV Methods of Analysis for Fluconazole in
Pharmaceutical Solid Dosage Forms. International
Journal of ChemTech Research. 1(4); 2009: 1131-1136.
19.
Araujo BV. et al. Validation of Rapid and Simple
LC–MS/MS Method for Determination of Voriconazole in Rat Plasma, Journal of Pharmaceutical and Biomedical
Analysis. 44; 2007: 985–990.
20.
Jennifer MC. et al. A High-Throughput LC–MS/MS
Method for the Quantitation of Posaconazole in Human Plasma: Implementing Fused
Core Silica Liquid Chromatography. Journal
of Pharmaceutical and Biomedical Analysis. 50; 2009: 46–52.
21.
Yu Y. et al. Development of a Simple Liquid
Chromatography-Tandem Mass Spectrometry Method for Multiresidue Determination
of Antifungal Drugs in Chicken Tissues. Journal
of AOAC International. 94 (5); 2011: 1650-1658.
22.
Sun N. et al. An Ultra-fast LC Method for the
Determination of Iodiconazole in Microdialysis Samples and its Application in
the Calibration of Laboratory-Made Linear Probes. Journal of Pharmaceutical and Biomedical
Analysis. 51(1); 2010: 248-51.
23.
Reddy NK. et al. Development and Validation of a
Stability Indicating UPLC Method for Determination of Voriconazole in
Pharmaceutical Formulation. Der
Pharmacia Letter. 3 (5); 2011: 249-259.
24. Laurent AD, et
al. Ultra-Performance Liquid Chromatography–Tandem Mass Spectrometry Method
for Simultaneous Quantification in Human Plasma of Fluconazole, Itraconazole,
Hydroxyitraconazole, Posaconazole, Voriconazole, Voriconazole-N-Oxide,
Anidulafungin, and Caspofungin. Antimicrobial
Agents and Chemotherapy. 54(12); 2010: 5303–5315.
25. Cheng SC,
Huang MZ, Shiea J. Thin Layer Chromatography/ Mass Spectrometry. Journal of Chromatography A .
1218(19); 2011: 2700-11.
Received on
07.12.2013 Modified on 15.12.2013
Accepted on
12.01.2014 © AJRC All right
reserved
Asian J. Research
Chem. 7(2): February
2014; Page 166-170