Development and Validation of RP-HPLC Method of Gabapentin in Pure and Pharmaceutical Formulations
Dilip Chandak, Pushpendra Sharma
Department of Chemistry, Sri Satya Sai University of Technology & Medical Sciences, Sehore-466001, Madhya Pradesh, India.
*Corresponding Author E-mail:
ABSTRACT:
A validated simple, sensitive, specific and precise RP-HPLC method was developed for the determination of Gabapentin in pure and pharmaceutical formulations. Method was carried on Zodiac C18 column (250 mm × 4.6 mm × 5µ particle size) using Methanol: Acetonitrile: Triethylamine (50: 25: 25) as mobile phase. Detection was carried out by U.V. at 211nm. The proposed method obeyed linearity in the range of 10-60µg/mL. and met all specifications as per ICH guidelines. Statistical analysis revealed that this method can be used in routine quality control studies of Gabapentin in pure and its formulations.
Instrumentation
To develop a High Pressure Liquid Chromatographic method for quantitative estimation of Gabapentin an isocratic PEAK HPLC instrument with Zodiac C18 column (250 mm x 4.6 mm, 5μ) was used.
The instrument is equipped with a LC 20AT pump for solvent delivery and variable wavelength programmable LC-7000 UV-detector. A 20μL Rheodyne inject port was used for injecting the samples. Data was analyzed by using PEAK software.
Standards and Chemicals Used:
Gabapentin was provided by Dr. Reddy’s laboratories. All the chemicals Acetonitrile, methanol, water, Triethylamine were HPLC grade, Merck Specialties Private Limited, Mumbai, India. Commercial tablets of Gabapentin were purchased from local market.
Preparation of the Mobile Phase:
Into a 1000mL cleaned volumetric flask, HPLC grade methanol 500mL, acetonitrile 250mL and Triethylamine 250 mL (which are filtered through 0.25mm membrane filters by vacuum filtration) were slowly added, mixed well and sonicated upto 20 min. Cool the above solution and pH was adjusted to 5.8 with orthophosphoric acid. This solution is again sonicated to 10 min. Cool the solution to room temperature and use for chromatography method.
Preparation of Standard Drug Solutions:
100mg of Gabapentin was accurately weighed and is dissolved in few mL of the mobile phase and sonicated for few min to dissolve the drug completely. Then it is filtered through 0.2μ ultipore filter paper and the volume is made upto 100mL with mobile phase to get a concentration of 1 mg/mL (free base) stock solution. This solution is further diluted with same solvent to obtain required working standard concentrations.
Sample Preparation:
20 commercial tablets of Gabapentin were finely powdered and the powder equivalent to 50mg of gabapentin accurately weighed to 50mL volumetric flask and dissolved in few mL of mobile phase. The above solution was subjected to sonication for 15min. after getting clear solution it is filtered through 0.25μm membrane filters and the solution is made upto 50mL with mobile phase resulting in preparation of 1mg/mL solution. This is further diluted so as to obtain required concentration of Gabapentin pharmaceutical dosage form.
METHODOLOGY:
The HPLC system was stabilized for thirty min. by passing mobile phase, detector was set at 211nm, flow rate of 1.0mL/min to get a stable base line. One blank followed by six replicates of a single standard solution was injected to check the system suitability. Six replicates of each standard solutions 10, 20, 30, 40, 50 and 60μg/mL were injected. Calibration graph was plotted by concentration of Gabapentin on X-axis and peak area on Y-axis. The amount of drug present in sample was computed in calibration graph.
Pharmaceutical Formulations:
Prepared dilution of pharmaceutical formulation is injected and the procedure described under bulk samples was followed. The amount of drug present in sample was computed in calibration graph.
Table 1: Optimized chromatographic conditions for estimation of Gabapentin
Parameter |
Condition |
Mobile phase |
Methanol: ACN: TEA (50 : 25 : 25) (v/v/v) |
Type of elution |
Isocratic |
pH |
5.8 |
Detection wavelength |
By U.V. at 211 nm |
Column |
Zodiac C18 column (250 x 4.6 mm, 5 μ) |
Column temp. |
Ambient |
Injection volume |
20 μL |
Flow rate |
1.0 mL/min. |
Run time |
10 min. |
Retention time |
4.58 min. |
RESULTS AND DISCUSSION:
The objective of the present work was to develop simple, precise and reliable HPLC method for the analysis of Gabapentin in bulk and pharmaceutical dosage form. This is achieved by using the most commonly employed column C18 with U.V. detection at 211nm. The representative chromatogram indicating Gabapentin is shown in Fig. 1.
ID |
Name |
Retain T |
Height |
Area |
Conc. |
Tail factor |
Theo. plate |
1 |
Gabapentin Standard Sum |
4.583 |
33658 |
244676.3 |
100.000 |
1.31 |
7923 |
|
244676.3 |
100.000 |
|
|
Fig. 1: Chromatogram of Gabapentin
Parameter fixation:
In developing this method, a systemic study of effects of various parameters was undertaken by varying one parameter at a time and controlling all other parameters. The following studies were conducted for this purpose.
Stationary phase characteristics:
Based on nature and solubility characteristics of gabapentin, reverse phase mode of HPLC was selected for chromatography. Among different RP-HPLC stationary phases tried C18 column was found to be optimum.
Mobile Phase Characteristics:
In order to get sharp peak with base line separation from interfering peaks carried out a number of experiments by varying the composition of solvents and mobile phase flow rate. To have an ideal separation of the drug under isocratic conditions, mixtures of solvents like methanol, water and acetonitrile with or without different buffers in different combinations were tested as mobile phase. A mixture of Methanol: ACN: TEA (50: 25: 25) (v/v/v) was proved to be the most suitable of all the combinations, since the chromatographic peak obtained was better defined and resolved and almost free from tailing.
Validation of the Proposed Method:
As an integral part of analytical method development is validation. The proposed method was validated as per ICH guidelines.
Linearity:
Linearity was performed by preparing standard solutions of Gabapentin at different concentration levels, twenty micro liters of each concentration was injected in into the HPLC system. The peak responses were read at 211nm and the corresponding chromatograms were recorded. Linearity plots of concentration over areas were constructed individually. Linearity results were obtained in the concentration range of 10-60μg/mL. Method of least square analysis was carried out for getting the slope, intercept and correlation coefficient values and the results were presented in Table 2.
Table 2: Linearity results of Gabapentin
Concentration (μg/mL) |
Peak area |
10 |
91926 |
20 |
172855 |
30 |
244676 |
40 |
317468 |
50 |
399191 |
60 |
481152 |
Precision:
Precision is the degree of repeatability of an analytical method under normal operational conditions. Precision of the method was performed as intraday precision, Inter day precision.
Intra-day precision:
To study the intra-day precision, six replicate standard solutions (30 μg/mL) of Gabapentin was injected. The percent relative standard deviation (% RSD) was calculated and it was found to be 0.96, which are well within the acceptable criteria of not more than 2.0. Results of system precision studies are shown in Table 3.
Table 3: Intraday Precision Values
Sample
|
Conc. (in ppm) |
Injection No.
|
Peak areas
|
RSD (Acceptance criteria ≤ 2.0%) |
Gabapentin |
20 |
1 |
244676 |
0.96 |
2 |
241923 |
|||
3 |
246782 |
|||
4 |
243921 |
|||
5 |
242016 |
|||
6 |
246296 |
Inter day precision:
To study the inter day precision, six replicate standard solutions (30 ppm) of Gabapentin was injected on third day of sample preparation. The percent relative standard deviation (% RSD) was calculated and it was found to be 0.811, which are well within the acceptable criteria of not more than 2.0.
Specificity:
The effect of wide range of excipients and other additives usually present in the formulation of Gabapentin in the determinations under optimum conditions were investigated in fact may have no observation at this UV Maximum. Chromatographic parameters maintained are specific for gabapentin
Ruggedness:
Percent recoveries of gabapentin was good under most conditions and didn’t show any significant change, when the critical parameters were modified.
Table 4: Ruggedness values
Sample
|
Conc. (in ppm)
|
Injection No.
|
Peak areas
|
RSD (Acceptance criteria ≤ 2.0%) |
Gabapentin |
20
|
1
|
255276 |
1.53 |
2
|
257254 |
|||
3 |
251972 |
|||
4 |
253284 |
|||
5 |
258697 |
|||
6
|
256061 |
Table 5: Limit of detection and limit of quantification for Gabapentin
Parameter |
Values |
Limit of quantification |
2.5 μg/mL |
Limit of detection |
0.75 μg/mL |
Table 6: Recovery Results
Level |
Concentration (μg/mL)
|
Amount of Gabapentin spiked (μg/mL) |
Total in ppm
|
Amount ofGabapentin recovered (ppm) |
% Recovery |
50% |
20 |
10 |
30 |
29.68881 |
98.96271 |
20 |
10 |
30 |
30.45844 |
101.5281 |
|
20 |
10 |
30 |
29.88266 |
99.60887 |
|
100% |
20 |
20 |
40 |
39.71273 |
99.28182 |
20 |
20 |
40 |
40.18874 |
100.4719 |
|
20 |
20 |
40 |
39.94784 |
99.86959 |
|
150% |
20 |
30 |
50 |
50.57591 |
101.1518 |
20 |
30 |
50 |
50.75891 |
101.5178 |
|
20 |
30 |
50 |
50.70931 |
101.4186 |
Accuracy:
The accuracy of the method was determined by standard addition method. A known amount of standard drug was added to the fixed amount of pre-analyzed standard solution. The standard addition method was performed at 50%, 100% and 150% level of 20 ppm. The solutions were analyzed in triplicate at each level as per the proposed method. The percent recovery and % RSD was calculated and results are presented in Table 6. Satisfactory recoveries ranging from 98% to 102% were obtained by the proposed method. This indicates that the proposed method was accurate.
Robustness:
The robustness study was performed by slight modification in flow rate of the mobile phase, pH of the buffer and composition of the mobile phase. Gabapentin at 2 ppm concentration was analyzed under these changed experimental conditions. It was observed that there were no marked changes in chromatograms, which demonstrated that the developed method was robust in nature.
Formulation:
S. No. |
Brand Name |
Concentration |
Area |
Amount found |
% Assay |
1 |
Neurontin -600 mg |
30 ppm |
241933 |
29.66
|
98.87 |
REFERENCES:
1. The Martindale 35th Edition, The Complete Drug Reference, Lambeta High Street, London (2006) pp. 362-363.
2. United States of Pharmacopoeia, 30, NF25, 32(6) (2007) p. 1689.
3. H. E. Abdellatef and H. M. Khalil, Colorimetric Determination of Gabapentin in Pharmaceutical Formulation, J. Pharma. and Biomed. Anal, 31(1), 209-214 (2003).
4. A. B. Ciavarella, A. Vilayat et al., Development and Application of a Validated HPLC Method for the Determination of Gabapentin and its Major Degradation Impurity in Drug Products, J. Pharma. and Biomed. Anal., 43, 1647-1653 (2007).
5. A. Ojha, R. Rathod, C. Patel and H. Padh, LC-MS Determination of Gabapentin from Human Plasma, Chromatographia, 66, 853-857 (2007).
6. M. Nomura, Y. Sakamoto, N. Nagashima, N. Takehara and K. Kitamura, Determination of Methylcobalamin in Serum by Liquid Chromatography-mass Spectrometry with Selected Ion Monitoring, Boil. Psychiatry, 42, 43 (1997).
7. M. Pujadas, S. Pichini, Civite, E. Santamarina, K. Perez, Rafel D. Torre, A Simple and Reliable Procedure for the Determination of the Psychoactive Drugs in Oral Fluid by Gaschromatography-Mass Spectrometry, J. Pharma. and Biomed. Anal., 4, 594-601 (2007).
8. B. R. Bharat, A. Shantha, P. Malairajan and S. Anbazhagan, Simulatneous Estimation and Validation of Gabapentin and Methylcobalmin in Combined Dosage form by RPHPLC, Indian Drugs, 44(10), 784-788 (2007).
9. R. T. Sane, U. Pendse, A. Moghe, S. Khedkar and P. Patil, Gabapentin in Pharmaceutical Preparation by HPTLC, Indian Drugs, 40(9), 547-548 (2003).
10. Varsha R. Galande, K. G. Baheti and M. H. Dehgham, UV-Visible Spectrophotometric Method for Estimation of Gabapentin and Methyl Cobalamin in Bulk and Tablet, Int. J. Chem. Tech. Res., 2(1), 695-699 (2010).